Dissolved Oxygen Meter With Extended Range - Ms 0257

 
INTRODUCTION
Ms 0257 is a water-resistant, dissolved oxygen meter that measures up to 300% saturation or 45 ppm (mg/L) with temperature compensation and automatic calibration. It has been developed for DO and temperature measurement in water, wastewater, and applications such as fish farming.
This instrument also allows altitude compensation up to 4000 m and the ppm and % saturation are both compensated for changes in solubility of oxygen in water and for permeability of the membrane as well as the temperature effect.
The included polarographic probe features built-in temperature compensation and removable protective membrane cover.
 
Dissolved Oxygen Meter With Extended Range - Ms 0257 Microsetcontrols
 
 
FEATURE
• On-screen tutorial messages
• Two point calibration
• Auto endpoint
• DO range up to 300%, temperature compensated
• Altitude compensation up to 4000 m
• Salinity compensation up to 80 g/L
• Automatic calibration in air
• GLP features
• Low battery indicator
 
 
 
 
Specification
Range         O2
:
0.00 to 45.00 mg/L (ppm)
Range          % Saturation O2
:
0.0 to 300.0%
Range         Temperature
:
0.0 to 50.0°C
Resolution         O2
:
0.01 mg/L (ppm)
Resolution         % Saturation O2
:
0.1%
Resolution Temperature
:
0.1°C
Accuracy         O2
:
±1.5% FS
Accuracy         % Saturation O2
:
±1.5% FS
Accuracy         Temperature
:
±0.2°C (excluding probe error)
Dissolved Oxygen Calibration
:
one or two points at 0%(zero solution) and 100% (in air)
Temperature Compensation
:
automatic, 0 to 50°C (32 to 122°F)
Altitude Compensation
:
0 to 4000 m (resolution 100 m)
Salinity Compensation
:
0 to 80 g/L (ppt) (resolution 1g/L)
Probe
:
Ms 87518/4F polarographic DO probe, internal temperature
sensor, DIN connector and 4 m cable (included)
Battery Type / Life
:
1.5V AAA (3) / approx 200 hours of continuous use without
backlight (50 hours with backlight on)
Environment
:
0 to 50°C (32 to 122°F); RH max 95%
Dimensions
:
185 x 72 x 36 mm (7.3 x 2.8 x 1.4")
Weight
:
300 g (10.6 oz)
 
 
request Microsetcontrols
button Microsetcontrols
 
line Microsetcontrols
           
   
Microsetcontrols
Microsetcontrols
Microsetcontrols  
   
 
Home

About us

Contact us

Enquiry

Sitemap
 
 
     
Industrial pH Sensor Ms pH 01 / Ms pH 02 / Ms pH 04, Ms pH 07, • pH Sensor – Industrial – HF Series Ms pH 22 / HF / Pt 100, • Industrial pH Sensor - Ms pH 36 – Flat (Origin USA), • pH Indicator Ms pH 19, • pH Controller Ms pH 27 / Ms pH 27 W, • pH Transmitter Ms pH 27 B / Ms pH 18 T, • Lab / Bench Meter Ms pH 616 / Ms pH 644, Ms 324 / Ms pH 816 / Ms 9535, • pH Sensor Housing for Fermentor Model 330, • pH Flow Cell, • Industrial ORP Sensor with Platinum Tip Ms ORP 01 / Ms ORP 02 / Ms ORP 36, • ORP Indicator Ms ORP 19, • ORP Controller Ms ORP 27 Panel Mounted, • ORP Controller Ms ORP 27 W Field Mounted
 
MICROSET CONTROLS
Microsetcontrols
:
B 404, Marvel Imperial, FP No 260/2&3, Narangibaug Road, Boat Club Area,
Pune - 411 001. Maharashtra, INDIA.
Microsetcontrols
:
Microsetcontrols
:


MICROSET CONTROLS
 

Microsetcontrols